Conductivity Meter and Cell

Contacting and Inductive Conductivity Meters and Cells

Simple, Flexible, Affordable

LAR LLC offers both contacting and inductive (torroidal) conductivity meters and probes in a range of cell constants and a variety of wetted materials to serve most municipal and industrial needs. Use the charts below to identify the sensor and controller that matches your requirements.

 

  • Contacting
  • Temp. Compensated
  • Industrial
  • 4-Electrode
  • Inductive
  • Accessories

Contacting Conductivity

 

down arrow  4222 Conductivity Meter for Contacting Cells

Conductivity ControllerPN: 9710610010
4222 Conductivity and Temperature Meter. Wall mounting 144x144
For Conductivity cell only
Measuring ranges: 00.00 - 20.0uS / 200.0uS / 2000uS / 20.00mS (200.0mS on demand).
Resolution 0.01/0.1/1uS / 00.01mS. Accuracy ± 0.2% f.s.
Temperature: - 10,0°C /130,0°C; resolution 1°C; accuracy 1% f.s
Automatic or manual temperature compensation.
Graphic display to visualize simultaneously: measurement, temperature, analog and digital outputs (set point) condition, alarms. Internal data logger (16,000 records). Tabular or graphic stored data display with maximum, minimum and average values. Programming key pad with 5 keys. Function keys to direct access to the calibration menu and to the graphs of the stored measure.
PID regulation to be set for analog or digital output (PWM or Frequency ).
Two digital Outputs to be set for set point, with programming of working range and activation time.
Digital output for Alarm: min/max, delay of the set point, residence time (live check), malfunction.
Digital output to be set for: Temperature set point or to control the rinsing cycle of the electrode with time and interval programming.
Digital input to be set for: external command of electrode rinsing cycle or for dosing inhibition
0/4 to 20mA primary output with limits that can be programmed within the measurement range.
0/4 to 20mA secondary output that can be programmed among: temperature/meas. repetition/PID function.
RS485 serial output with MODBUS RTU protocol.
Able to simulate the outputs via a keyboard.

Hardware technical features:
Graphic display LCD STN 128x64 back-lighted. 5 keys programming keyboard.
n. 1Input for Temperature sensor PT 100 with 3 or 4 wires (std) NTC or Pt 1000
Data logger with Flash memory 4 Mbit ( 16.000 records), Type: f.i.f.o. or filling
n. 2 analog outputs 0/4÷20mA 500Ohm galvanically isolated
n. 4 digital outputs– Dry contact Relay
n. 1 digital input - Active and already supplied
n. 1 RS 485 serial output MODBUS protocol
Power supply 100-240Vac 47÷63Hz
ABS container for panel mounting. Mechanical protection IP66 Dim. mm. 144x144x122,5 (p) Weight Kg. 0.9

9710610000
4222 Conductivity and Temperature Analyzer. Panel mounting 144x144
Same technical features as for 4222. Grey ABS case for panel mounting.
Dim. mm 144x144x61,0 (p) Weight Kg. 0.9

PN: 9710610050
4222 Conductivity and Temperature Analyzer. Panel mounting 96 x96
Same technical features as for 4222 but LCD STN 57x30 back-lighted graphic display.
Black ABS case for panel mounting. Mechanical Protection IP 20. Dim. mm 96x96x123 (p) Weight Kg. 0.9

PN: 9400970000
Internal USB port to download stored data via USB flash memory

PN: 9400970001
Internal Supply board 24V ac/dc

 

down arrow Analyzer icon 42XX-22 Multi-parameter Conductivity Meter

PN: 9720610010
4238-22 Combined Meter for Conductivity + pH/Rx + Temperature. Wall mounting 144x144

PN: 9720610000
4238-22 Combined Meter for Conductivity + pH/Rx + Temperature. Panel mounting 144x144

PN: 9720610050
4238-22 Combined Meter for Conductivity + pH/Rx + Temperature. Panel mounting 96 x 96

PN: 9720615010
4293-22 Combined Meter for Conductivity + Chlorine/ClO2/O3 + Temp. Wall mount 144x144

PN: 9720615000
4293-22 Combined Meter for Conductivity + Chlorine/ClO2/O3 + Temp. Panel mount 144x144

PN: 9720615050
4293-22 Combined Meter for Conductivity + Chlorine/ClO2/O3 + Temp. Panel mounting 96 x 96

PN: 9720619014
4283-22 Combined Meter for Conductivity + Oxygen + Temp. Wall mount 144x144
For Optical D.O. sensor S423/N/OPT only

PN: 9720619004
4283-22 Combined Meterfor Conductivity + Oxygen + Temp. Panel mount 144x144
For Optical D.O. sensor S423/N/OPT only

PN: 9720619054
4283-22 Combined Meter for Conductivity + Oxygen + Temp. Panel mount 96 x 96
For Optical D.O. sensor S423/N/OPT only

 

down arrow Analyzer icon 3522 Conductivity and Temperature Meter for Contacting Cells

PN: 9700611010

3522 Conductivity and Temperature Meter. Wall mounting 144x144
For Conductivity cell only
Measuring ranges 00.00 - 20.00uS / 200.0uS / 2000uS / 20.00mS.
Resolution 0.01/0.1/1uS / 00.01mS. Accuracy ± 0.2% f.s.
Temperature: - 10.0°C /130,0°C; resolution 1°C; accuracy 1% f.s
Automatic or manual temperature compensation.
Graphic display to visualise simultaneously: measurement, temperature, analog and digital outputs condition, alarms. Programming key pad with 4 keys. Function key to direct access to calibration menu.
PID regulation to be set on digital output (PWM or Frequency ).
Two digital Outputs to be set for set point, with programming of working range (hysteresis/direction) as well as of the activation time.
Digital output for Alarm: minimum, maximum, delay of the set point, residence time (live check), malfunction.
Digital output to control the rinsing cycle of the electrode with time and interval programming.
Digital input to be set for: external command of electrode rinsing cycle or for dosing inhibition
0/4 to 20mA primary output with programmable limits within the measurement range.
Able to simulate the outputs via a keyboard.

Hardware technical features:

Graphic display LCD STN 128x64 back-lighted. 4 keys programming keyboard.
n. 1Input for Temperature sensor PT 100 with 3 or 4 wires (std) NTC or Pt 1000.
n. 1 analog output 0/4÷20mA 500Ohm galvanically isolated
n. 4 digital outputs– Dry contact Relay
n. 1 digital input - Active and already supplied
n. 1 RS 485 serial output
Power supply 100-240Vac 47 to 63Hz
ABS container for panel mounting. Mechanical protection IP66 Dim. mm. 144x144x122,5 (p) Weight Kg. 0.9

PN: 9700611000

3522 Conductivity & Temperature Meter. Panel mounting 144x144
Same technical features as for 3522. Grey ABS case for panel mounting.
Dim. mm 144x144x61,0 (p) Weight Kg. 0,9

PN: 9700611050

3522 Conductivity & Temperature Meter. Panel mount 96 x96mm
Same technical features as 3522 but LCD STN 57x30 back-lighted graphic display.
Black ABS case for panel mounting. Mechanical Protection IP 20. Dim. mm 96x96x123 (p) Weight Kg. 0.9

 

down arrowS411S, K=1, Conductivity Cell, Measuring Range 0 to 2.000 µS, No Cable

PN: 6100011100

S411S conductivity cell

PVC body 1-inch threaded.
Stainless steel electrodes.
Constant K=1.
Measuring range 0 to 2.000 µS
Max working temperature 50°C
Max working pressure 1 bar.
Without cable

 

down arrowS411, K=1, Conductivity Cell, Measuring Range 0 to 20.000 µS, 5M Cable

PN: 6100010140

S411 conductivity cell

PVC body
1-inch threaded Graphite electrodes
Constant K=1.
Measuring range 0 to 20.000 µS
Max working temperature 50°C
Max working pressure 2 bar
5m fixed cable

 

down arrowS411, K=1, Conductivity Cell, Measuring Range 0 to 20.000 µS, 10M Cable

PN: 6100010145
S411 conductivity cell
PVC body
1-inch threaded Graphite electrodes
Constant K=1.
Measuring range 0 to 20.000 µS
Max working temperature 50°C
Max working pressure 2 bar
10m fixed cable

 

down arrowS411/TEF, K=1, Conductivity Cell, 0 to 10.000 µS, 5M Cable

PN: 6100010340
S411/TEF conductivity cell
PTFE body
1" threaded Stainless steel electrodes.
Constant K=1.
Measuring range 0 to 10.000 µS.
Max working temperature 100°C
Max working pressure 2 bar.
5m fixed cable

Temperature Compensated Conductivity Cells

 

down arrowS411/C, K=1, Temp-compensated Cell, 0 to 20.000 µS, 5m cable

PN: 6100010146
S411/C PT 100 Thermo-compensated (PT100) conductivity cell
PVC body 1" threaded Graphite electrodes.
Constant K=1.
Measuring range 0 to 20.000 µS
Max working temperature 50°C.
Max working pressure 2 bar
5m fixed cable

 

down arrowS411/C, K=1, Temp-compensated Cell, 0 to 20.000 µS, 10m cable

PN: 6100010147
S411/C PT100 Thermo-compensated (PT100) conductivity cell
PVC body 1" threaded Graphite electrodes.
Constant K=1. Measuring range 0 to 20.000 µS
Max working temperature 50°C.
Max working pressure 2 bar
10m fixed cable

 

down arrowS411/TEF/C, K=1, Temp-compensated Cell, 0 to 10.000 µS, 5m cable

PN: 6100010342
S411/TEF/C PT100 Thermo-compensated (PT100) conductivity cell
PTFE body 1" threaded
Stainless steel electrodes.
Constant K=1.
Measuring range 0 to 10.000 µS.
Max working temperature 100°C
Max working pressure 2 bar.
5m fixed cable


Conductivity Cells for Industrial Applications

down arrowS428, K=0.1, Glass/Platinum conductivity cell, Range: 2 to 100 mS

PN: 6110000140
S428 Glass/Platinum conductivity cell
High measuring range K=0.1
Glass body with 3 Platinum electrodes.
Measuring range: 2 to 100 mS
Max working temperature 100°C
Max working pressure 0.2 bar.
Dimension mm. 12 x 120 ( Øxl)
5m fixed cable

 

down arrowS411/U K=0.1, Temp Compensated conductivity cell, 10 to 200 mS

PN: 6100002640
S411/U K0.1 Temp Compensated, conductivity cell
PES body with Graphite electrodes.
threaded connection to the process ½"NPT.
Constant K=0.1
Measuring range 10 to 200 mS.
Max working temperature 120°C
Max working pressure 6 bar at 20°/2 bar at 120°C.
5 mt. cable with water tight connector (other lengths on request)

 

down arrowS411/U K=0.1, Temp Compensated conductivity cell, 10 to 200 mS, no cable

PN: 6100002600
S411/U K0.1 Temp Compensated, conductivity cell
PES body with Graphite electrodes.
Threaded connection to the process 1/2-inch NPT.
Constant K=0.1
Measuring range 10 to 200 mS.
Max working temperature 120°C
Max working pressure 6 bar at 20°/2 bar at 120°C.
No cable

 

down arrowS411/U K1 Temp Compensated conductivity cell, K=1, 0 to 50.000µS, 5M cable

PN: 6100012640
S411/U K1 Temp Compensated conductivity cell
PES body with Graphite electrodes
threaded connection to the process 1/2-inch NPT.
Constant K=1
Measuring range 0 to 50.000µS
Max working temperature 120°C
Max working pressure 6 bar at 20°C, 2 bar at 120°C
5M cable with water tight connector (other lengths on request)

 

down arrowS411/U K1 Temp Compensated conductivity cell, 0 to 50.000µS, K=1, no cable

PN: 6100012600
S411/U K1 Temp Compensated conductivity cell
PES body with Graphite electrodes
threaded connection to the process ½"NPT.
Constant K=1
Measuring range 0 to 50.000µS
Max working temperature 120°C
Max working pressure 6 bar at 20°/2 bar at 120°C
Without cable

 

down arrowS411/U K10 Temp Compensated conductivity cell, 0.05 to 200µS, K=0.1, 5M cable

PN: 6100022640
S411/U K10 Temp Compensated conductivity cell
PES body with Graphite electrodes.
threaded connection to the process ½" NPT.
Constant K=0.1
Measuring range 0.05 to 200µS
Max working temperature 120°C
Max working pressure 6 bar at 20°C, 2 bar at 120°C
5M cable with water tight connector (other lengths on request)

 

down arrowS411/U, K=0.1, Temp Compensated conductivity cell, 0.05 to 200µS, no cable

PN: 6100022600
S411/U K10 Temp Compensated conductivity cell
PES body with Graphite electrodes.
threaded connection to the process 1/2-inch NPT.
Constant K=0.1
Measuring range 0.05 to 200µS
Max working temperature 120°C
Max working pressure 6 bar at 20°C, 2 bar at 120°C
No cable

 

down arrowS411/P K=100, Temp Compensated conductivity cell, 0.04 to 20 µS, 5M Cable

PN: 6100032240
S411/P K100 Temp Compensated conductivity cell
SS AISI 316 body and electrodes
threaded connection to the process 1/2-inch NPT.
Constant K=100 Measuring range 0.04 to 20 µS
Max working temperature 130°C
Max working pressure 16 bar at 130°C
5M cable with water tight connector (other lengths on request)

 

down arrowS411/P, K=100, Temp Compensated conductivity cell, 0.04 to 20 µS, No Cable

PN: 6100032200
S411/P K100 Temp Compensated conductivity cell
SS AISI 316 body and electrodes
threaded connection to the process 1/2-inch NPT.
Constant K=100 Measuring range 0.04 to 20 µS
Max working temperature 130°C
Max working pressure 16 bar at 130°C
Without cable

 

down arrowS411/P K10, K=100, Temp Compensated conductivity cell, 0 to 1000 µS, 5M Cable

PN: 6100022240
S411/P K 10 Temp Compensated conductivity cell
SS AISI 316 body and electrodes.
threaded connection to the process ½"NPT.
Constant K=100
Measuring range 0 to 1000 µS
Max working temperature 130°C
Max working pressure 16 bar at 130°C
5 mt. cable with water tight connector (other lengths on request)

 

down arrowS411/P K10, K=100, Temp Compensated conductivity cell, 0 to 1000 µS, No Cable

PN: 6100022200
S411/P K 10 Temp Compensated conductivity cell
SS AISI 316 body and electrodes.
threaded connection to the process ½"NPT.
Constant K=100
Measuring range 0 to 1000 µS
Max working temperature 130°C
Max working pressure 16 bar at 130°C
Without cable

 

4-Electrode Cell, Wide Range, Immersion- or insertion-Style

down arrowS411/4E, K=0.55, Temp Compensated Cell, 0 to 1000mS (1 Siemens), 5M Cable

PN: 6120010500
S411/4E Temp Compensated conductivity cell K=0.55
Epoxy body and graphite electrodes (4). Measuring range 0 to 1000mS (1 Siemens)
threaded connection to the process 13.5 PG.
Max working temperature 70°C Max working pressure 4 bar at ambient temperature
5m fixed cable

 

Inductive (Torroidal) Conductivity

 

down arrow analyzer-icon 4223 Conductivity and Temperature Meter

PN: 9710611010
4223 Conductivity and Temperature Analyzer. Wall mounting 144x144
For inductive cell only Measuring ranges: 00.00 to 2000µS / 20.00 / 200.00mS / 2Siemens
Automatic or manual temperature compensation. Accuracy ± 0.2% f.s.
Graphic display to visualize simultaneously: measurement, temperature, analog and digital outputs (set point) condition, alarms. Internal data logger (16,000 records). Tabular or graphic stored data display with maximum, minimum and average values. Programming key pad with 5 keys. Function keys for access to calibration menu and graphs of stored measurements.
PID regulation to be set for analog or digital output (PWM or Frequency).
Two digital Outputs for set point, with programming of working range and activation time.
Digital output for Alarm: min/max, setpoint delay, residence time (live check), malfunction.
Digital output to be set for: Temperature set point or to control the rinsing cycle of the electrode with time and interval programming.
Digital input to be set for: external command of electrode rinsing cycle or for dosing inhibition
0/4 to 20mA primary output with limits that can be programmed within the measurement range.
0/4 to 20mA secondary output. programmable for: temperature/meas. repetition/PID function.
RS485 serial output with MODBUS RTU protocol.
Ability to simulate outputs via keyboard.
Hardware technical features:
Graphic display LCD STN 128x64 back-lighted. 5 keys programming keyboard.
n. 1 Input for Temperature sensor PT 100 with 3 or 4 wires (std) NTC or Pt 1000
Data logger with Flash memory 4 Mbit ( 16.000 records), Type: f.i.f.o. or filling
n. 2 analog outputs 0/4 to 20mA 500 Ohm galvanically isolated
n. 4 digital outputs– Dry contact Relay
n. 1 digital input - Active and already supplied
n. 1 RS 485 serial output MODBUS protocol
Power supply 100-240VAC 47 to 63Hz
ABS container for panel mount. Mechanical protection IP66 Dim. mm. 144x144x122.5 (p)
Weight Kg. 0.9

PN: 9710611000
4223 Conductivity and Temperature Analyzer. Panel mounting 144x144
For inductive cell only Measuring ranges: 00.00 to 2000µS / 20.00 / 200.00mS / 2Siemens
Automatic or manual temperature compensation. Accuracy ± 0.2% f.s.
Graphic display to visualize simultaneously: measurement, temperature, analog and digital outputs (set point) condition, alarms. Internal data logger (16,000 records). Tabular or graphic stored data display with maximum, minimum and average values. Programming key pad with 5 keys. Function keys for access to calibration menu and graphs of stored measurements.
PID regulation to be set for analog or digital output (PWM or Frequency).
Two digital Outputs for set point, with programming of working range and activation time.
Digital output for Alarm: min/max, setpoint delay, residence time (live check), malfunction.
Digital output to be set for: Temperature set point or to control the rinsing cycle of the electrode with time and interval programming.
Digital input to be set for: external command of electrode rinsing cycle or for dosing inhibition
0/4 to 20mA primary output with limits that can be programmed within the measurement range.
0/4 to 20mA secondary output. programmable for: temperature/meas. repetition/PID function.
RS485 serial output with MODBUS RTU protocol.
Ability to simulate outputs via keyboard.
Hardware technical features:
Graphic display LCD STN 128x64 back-lighted. 5 keys programming keyboard.
n. 1 Input for Temperature sensor PT 100 with 3 or 4 wires (std) NTC or Pt 1000
Data logger with Flash memory 4 Mbit ( 16.000 records), Type: f.i.f.o. or filling
n. 2 analog outputs 0/4 to 20mA 500 Ohm galvanically isolated
n. 4 digital outputs– Dry contact Relay
n. 1 digital input - Active and already supplied
n. 1 RS 485 serial output MODBUS protocol
Power supply 100-240VAC 47 to 63Hz
ABS container for panel mount. Mechanical protection IP66 Dim. mm. 144x144x122.5 (p)
Weight Kg. 0.9

PN: 9710611050
4223 Conductivity and Temperature Analyzer. Panel mounting 96 x96
Same technical features as for 4223 but LCD STN 57x30 back-lighted graphic display.
Black ABS case for panel mounting. Mechanical Protection IP 20. Dim. mm 96x96x123 (p) Weight Kg. 0.9

PN: 9400970000 Internal USB port to download stored data via USB flash memory

PN: 9400970001 Internal Supply board 24V ac/dc

 

down arrow analyzer-icon Multi-parameter Inductive Conductivity Meter

PN: 9720613010
4238-23 Combined Analyzer for Inductive Conductivity + pH/Rx + Temp. Wall mounting 144x144

PN: 9720613000
4238-23 Combined Analyzer for Inductive Conductivity + pH/Rx + Temp. Panel mounting 144x144

PN: 9720613050
4238-23 Combined Analyzer for Inductive Conductivity + pH/Rx + Temp. Panel mounting 96 x 96

PN: 9720616010
4293-23 Combined Analyzer for Inductive Conductivity + Cl/ClO2/O3 + Temp.. Wall mount 144x144

PN: 9720616000
4293-23 Combined Analyzer for Inductive Conductivity + Cl/ClO2/O3 + Temp. Panel mount 144x144

PN: 9720616050
4293-23 Combined Analyzer for Inductive Conductivity + Cl/ClO2/O3 + Temp. Panel mount 96 x 96

 

down arrow analyzer-icon 3523 Inductive Conductivity Meter

PN: 9700612010
3523 Conductivity and Temperature Analyzer. Wall mounting 144x144
For inductive cell only Measuring ranges: 00.00 to 2000µS /÷ 20.00/200,00mS/ 2Siemens
Resolution 0.01/0.1/1uS / 00.01mS. Accuracy ± 0.2% f.s.
Temperature: - 10.0°C /130.0°C; resolution 1°C; accuracy 1% f.s
Automatic or manual temperature compensation.
Graphic display to visualise simultaneously: measurement, temperature, analog and digital outputs condition, alarms. Programming key pad with 4 keys. Function key to direct access to calibration menu.
PID regulation to be set on digital output (PWM or Frequency ).
Two digital Outputs to be set for set point, with programming of working range (hysteresis/direction) as well as of the activation time.
Digital output for Alarm: minimum, maximum, delay of the set point, residence time (live check), malfunction.
Digital output to control the rinsing cycle of the electrode with time and interval programming.
Digital input to be set for: external command of electrode rinsing cycle or for dosing inhibition
0/4 to 20mA primary output with programmable limits within the measurement range.
Able to simulate the outputs via a keyboard.

Hardware technical features:

Graphic display LCD STN 128x64 back-lighted. 4 keys programming keyboard.
n. 1 Input for Temperature sensor PT 100 with 3 or 4 wires (std) NTC or Pt 1000.
n. 1 analog output 0/4÷20mA 500Ohm galvanically isolated
n. 4 digital outputs– Dry contact Relay
n. 1 digital input - Active and already supplied
n. 1 RS 485 serial output
Power supply 100-240 VAC 47 to 63Hz
ABS container for panel mounting. Mechanical protection IP66 Dim. mm. 144x144x122.5 (p) Weight Kg. 0.9

PN: 9700612000
3523 Conductivity and Temperature Analyzer. Panel mounting 144x144
Same technical features as 3523. ABS case for panel mounting. Dim. mm 144x144x61,0 (p) Weight Kg. 0,9

PN: 9700612050
3523 Conductivity and Temperature Analyzer. Panel mounting 96 x96
Same technical features as for 3523 but LCD STN 57x30 back-lighted graphic display.
Black ABS case for panel mounting. Mechanical Protection IP 20. Dim. mm 96x96x123 (p) Weight Kg. 0.9


down arrowS411/IND Temp. Compensated Inductive Cell

PN: 6100011441
S411/IND Thermocompensated Inductive cell
Glass Loaded Polypropylene body. Measuring range 0 to 1000mS (1 Siemens)
threaded connection to the process 0.5 BSP
Max working temperature 60°C Max working pressure 6.5 bar at ambient temperature
5m fixed cable

 

down arrowS411/IND/T/60 Temp. Compensated Inductive cell immersion style 600mm

PN: 6100011442
S411/IND/T/60 Thermocompensated Inductive cell immersion style 600mm
Technical features as S411/IND plus PVC immersion pipe L=600mm.

 

down arrowS411/IND/T/120 Temp. Compensated Inductive cell immersion style 1200mm

PN: 6100011443
S411/IND/T/120 Thermocompensated Inductive cell immersion style 1200mm
Technical features as S411/IND, plus PVC immersion pipe L=1200mm.

 

Accessories

 

down arrowFlanges, Tees, Probe Holders, Standard Solution, Cables

PN: 9400410091 Tank wall angled flange adapter for S411/IND/T

PN: 9400410092 Stainless steel mounting bracket for S411/IND/T

PN: 9400410093 1" PVC adjustable flange for S411/IND/T

PN: 6100011444 S411/IND/E/1.5 Thermocompensated Inline Inductive sensor supplied in a 1.5" plain PVC tee. Wetted materials glass filled polypropylene, PVC and Viton.
Maximum operating temperature 60 °C, Maximum pressure 6.5 Bar (100 psi)
5 metre connection cable with tag ends."

PN: 6100011445 S411/IND/E/0.5 Thermocompensated Inline Inductive sensor supplied in a 1.5" plain PVC tee. with 0.5" BSP Galvanised iron fittings.
Wetted materials glass filled polypropylene, PVC, galvanised mild steel and Viton
Maximum operating temperature 60°C, Maximum pressure 6.5 Bar (100 psi)
5 metre connection cable with tag ends.

PN: 6100011446 S411/IND/INS Thermocompensated Inline Inductive sensor 1.25" BSP Male threaded insertion. Maximum operating temperature 60 °C, Maximum pressure 6.5 Bar (100 psi)
5 metre connection cable with tag ends.
Complete with 5 metre connection cable.

Probe holder for S411S, S411, S411C,S411TEF,S411/TEF/C

PN: 9500430000 S315/C/50 Immersion probe holder
PP body. Complete with terminal flange for electrode protection Dim: ( Ø x l ) 63 x 570 mm."

PN: 9500430001 S315/C/100 Immersion probe holder
PP body. Complete with terminal flange for electrode protection Dim: ( Ø x l ) 63 x 1070 mm"

PN: 9500430002 S315/C/150 Immersion probe holder
PP body. Complete with terminal flange for electrode protection Dim: ( Ø x l ) 63 x 1570 mm"

PN: 7600430001 Nozzle/63 for automatic rinsing of electrode S315 Ø 63 mm. Stainless Steel body.

Standard Solution, cables accessories

PN: 7202020703 S608 1413µS Standard solution 1X500 ml. set

PN: 7202020704 S608 5µS Standard solution 1X300 ml. set DFM (eq SIT ) certificated

PN: 1620100400 S332 cable 4 wire + shield. For thermocompensated conductivity cell. Price x meter

PN: 1663400202 S332 cable 2 wire + shield. For non thermocompensated conductivity cell. Price x meter

PN: 7989000000 5 m. cable with water tight connector for S411/U and S411/P conductivity cell

PN: 7989000001 10 m.cable with water tight connector for S411/U and S411/P conductivity cel

PN: 7989000002 15 m. cable with water tight connector for S411/U and S411/P conductivity cell

PN: 8109000045 S321 Junction box IP 66 complete with terminal board